In Vitro Precision |
Coefficient of variation (CV) 5% |
Glucose Range |
36 - 450mg/dl (or 2-25mM/L) |
Response Time (sec) |
100 - 300s |
In vivo calibration |
In vivo calibration |
Interference species |
Acetaminophen, ascorbic acid, uric acid |
Length |
5cm |
Sensor Size |
0.6— 0.7mm |
Reference Electrode |
Ag/AgCl |
Polarization voltage (V) |
0.6 |
Sensor Life |
3-4 months in solutions at room temperature under continuous polarization; 15-30 days in vivo |
Shelf Life |
6 months |
Operation Conditions |
20° to 40°C (68° to 104°F) |
Storage Conditions |
10° to 25°C (50° to 77°F) |