BLPR2  Blood Pressure Transducer & Cable 
 
  • Used for direct arterial and venous pressure measurement
  • Accurate, linear, and stable with temperature
  • Compatible with TBM4M and BP1
 
 

Features

BLPR2 can be used for the direct arterial and venous pressure measurement in animal blood vessels. Supplied sterile, BLPR2 is accurate, linear and stable with temperature. May be sterilized cold with Cidex or a similar bactericide.

Cable and compatibility

BLPR2 is equipped with a twelve-foot cable and connector compatible with WPI’s four-channel signal conditioning unit, SYS-TBM4M Transbridge, and the single-channel BP1 blood pressure monitor. Cable has moisture-resistant locking connector. A continuous, uniform lumen eliminates places for bubbles to form and lodge. The clear fluid path is easy to inspect. Easy to mount - slotted transducer body accepts Velcro strap.

Easy set-up

To facilitate setup and operation, a four-way stopcock that allows easy filling, flushing, and zeroing of the transducer is included. Typically, the stopcock is located between the transducer and the animal catheter where it can be used to quickly zero, flush, or de-bubble the transducer. 

Specifications 

WORKING PRESSURE -50 to + 300mmHg
OVERPRESSURE -400 to +4000mmHg
EXCITATION VOLTAGE 1-10VDC or RMS to 5kHz
INPUT IMPEDANCE 350Ω ±10%
OUTPUT IMPEDANCE 300Ω ±10%
ISOLATION <5µA leakage at 120VAC
SENSITIVITY 5µV/V/mm Hg
DYNAMIC RESPONSE 100Hz
EIGHT HOUR DRIFT 1mmHg after 5 minute warm-up
MAXIMUM ERROR Total combined effects of Sensitivity, Linearity,
ENVIRONMENTAL: Operating Temp. Range +15°C to +40°C
ENVIRONMENTAL: Storage Temp. Range -30°C to +60°C
ENVIRONMENTAL: Light Sensitivity Not affected by light under 3000 ft/candles
SHIPPING WEIGHT 1 lb.

 
 
 